Proving an Intelis Gas Meter
Intelis Gas Meters support the ability to connect a prover to ensure the proper calibration of the meter. The prover parameters can be adjusted in Device Configuration node or in the file snapshot for the device.
Pulse weight. Pulse weight is the volume associated with an individual pulse measured in cubic feet or cubic decimeters. The default value is 0.10 cfh.
Pulse width. The length of time, measured in milliseconds, that a single pulse lasts. The default value is 15 milliseconds.
Maximum time. The maximum time the meter will remain in test mode before automatically returning to normal mode. The default value is one hour.